Envent Engineering Ltd – 131S & 132S Software Manual – Revision 1
Page 22
4.4
Streams
Streams Configuration
Allows the user to define up to 10 unique streams. These streams can be process
(customer), calibration, reference, or baseline streams. See the tables, below, for
a description of each of the sections on this page.
Stream Name
User-defined name (ex: “Sales Gas” or “Sales Cal”)
Stream Type
Process = Process (customer end) stream
Reference = Calibration standard as an “unknown”
Calibration = Calibration standard, can adjust calibration
factors
Baseline = Can run without sample
Purge Time
[Optional] time to purge the stream before running
Purge Outputs
[Optional] outputs which will be turned on to purge the
stream
Valid Run Count
[Calibration Only] controls the number of runs the
calibration will average for the final report
Components
User can select which components are used on the stream
(defined in Component Properties)
Normalize
Checked = Values are normalized to 100%
Unchecked = Values are not normalized