Envent Engineering Ltd – 131S & 132S Software Manual – Revision 1
Page 27
Activation
Allows the user to define which system events cause
the stream to run. The check box activates the event,
and the number next to the check box allows the
user to define how many runs the event will do.
The example on the left shows a typical calibration
configuration, where the manual calibration button
runs the stream 3 times. There is also an interval to
calibrate daily at 8 AM; again, running 3 times.
Telemetry
The telemetry section allows the user to define
telemetry for the stream. When the stream is
running, the selected discrete outputs will be active.
In the example to the left, the CAL-1 LED has been
checked which will illuminate the CAL-1 LED when
the calibration stream is running.
Lockouts
Lockouts allow the user to specify when to allow the
stream to run and when to lock it out. When the
Lockouts menu is clicked, the blue device register
menu opens on the right-side of the software.
An example of a low-helium lockout can be seen to
the left.