Envent Engineering Ltd – 131S & 132S Software Manual – Revision 1
Page 42
Analysis Sequence Dashboard
This view allows the user to see the sequence of analysis events to occur. The
elements on the left side of the screen are buttons with two distinct functions.
Clicking the body of the button (gray area) will queue up the stream in the
analysis sequence according to the “Manual” system event for that stream
(defined in the Components page of Configuration). Clicking the “Continuous”
checkbox will queue up the stream in the analysis sequence but will run
indefinitely until the checkbox is unchecked.
The “Abort Current Run” button (at the top) will immediately cancel the analysis,
regardless of the timing of the run, and put the analyzer into idle mode. Extreme
caution must be taken when using this button, as it can harm the operation of the
analyzer since components are at an unknown position in the analytical
columns/valves when the button is pressed.