11
Appendix
Entry offset
Function group
OFFSET
V
V2
Calibration status
is displayed
o.k.
E---
V3
Store
offset
no; new
yes;
V4
Entry of absolute value
current measured value
V1
Function group
CALIBRATION
C
Store calibration
result
non; new
yes;
C115
Entry of concentration
of first calibration solution
entire meas. range
100.0 FNU;
100.0 ppm (mg/l);
10.0 g/l; 10.0 %
.
C111
Entry of concentration
fo second cal. solution
last calibration value
C112
C111 × 1.1
≥
C112
Entry of concentration
of third calibration solution
last calibration value
C113
C112 × 1.1
≥
.
C113
Calibration status
is displayed
o.k.
E---
C114
Calibration selection
Data
= calibration data;
only C1 (6) is available
for data set 1
1-pt = 1-point
calibration
Refl
= reflection adaptation
Edit = edit
calibration
Corr
= 3-point correction
3-pt = 3-point
calibration
C1 (6)
C1 (5)
C1 (4)
C1 (3)
C1 (2)
C1 (1)
Store calibration
result
non; new
yes;
C135
Entry of concentration
of second cal. solution
> value in C131 × 11
current value in C112
Entry of concentration
of first calibration solution
entire measuring range
current value in C111
C132
C131
Entry of concentration
of third calibration solution
> value in C132 × 1.1
current value in C113
C133
Calibration status
is displayed
o.k.
E---
C134
Store calibration
results
; no; new
yes
C123
C121
C122
Entry of correct
concentration of
calibration solution 3
entire meas. range
current value in C113
Calibration status
is displayed
o.k.
E---
Entry of correct
measured value
0 ... 2.0 FNU
0 ... 5.0 ppm (mg/l)
0;
0;
C141
Calibration status
is displayed
o.k.
E---
C142
Store calibration
results
; no; new
yes
C143
Store calibration
results
; no; new
yes
C153
Entry of current
calibration value
entire measuring range
current measured value
C151
Calibration status
is displayed
o.k.
E---
C152
Display of cal. point 1
Comparative value
C161
Display of cal. point 2
Comparative value
C162
Display of cal. point 3
Comparative value
Display of
slope 1
current value
C163
C164
Display of
slope 2
current value
C165
MEAS. VALUE DISPLAY
with
TEMPERATURE DISPLAY
in °C
Measured value display
in FNU
Temperature display
suppressed
Temperature display
in °F
Other errors are
displayed
(up to 10 errors)
1st error is displayed
(if present)
Edit
mode:
Code
22
Read
mode:
an
y
code
Edit
mode:
Code
22
Read
mode:
an
y
code
Select alarm
delay unit
min
s;
F2
Alarm delay
0 ... 2000 s (min)
(depends on F2)
0s (min)
F3
Error current
setting
2.4 mA
22 mA
F4
Error number
selection
1 ... 255
1
F5
Set alarm contact
to be effective
no
yes;
F6
O231
Tablen option selection
edit
read
O232
Set number of
table value pairs
1 ... 10
1
O233
Selection of
table value pairs
1 ... number of
table value pairs;
asign
.
1
O234
Entry of x value
(measured value)
entire meas. range
0.00 FNU / ppm (mg/l)
g/l / %
Entry of simulation value
0 ... 22.00 mA
current value
O221
Function group
CURRENT OUTPUT
O
Currect output selection
; Out2
Out1
O1
Characteristic selection
Tab = table
sim = simulation
lin = linear
O2 (3)
O2 (2)
O2 (1)
0/4 mA value
entry
entire meas. range
0.0 FNU; 0.0 ppm (mg/l);
0.0 %; 0.0 °C
O212
Current range selection
0-20 mA
4-20 mA;
O211
20 mA value
entry
entire meas. range
9999 FNU; 9999 ppm(mg/l);
300.0 g/l; 10.0 %; 100.0 °C
O213
Function group
SETUP 2
B
Switch wiper control
on or off
on
off
B1
Set duration of
wiper operation
3 ... 999 s
30
B2
Set pause time between
wiper cycles
1 ... 7200 min
120 min
B3
Selection of calibration
data set used
1 ... 3
3
B4
Copy data sets
1
2; 1
3
2
3; 3
2
no
→
→
→
→
B5
Display of measured value
with reflection adaptation
; no
yes
B6
Function group
ALARM
F
Select contact type
= steady contact
Fleet = fleeting contact
Steady
F1
Function group
SETUP 1
A
Selection of oper. mode
; ppm; mg/l; g/l;
%; spec.
FNU
A1
Selection of unit
displayed (if A1=spec)
; t/m
%; none
kg/l
3
A2
Display of sensor
connected
; CUS 41
CUS 31
A4
Display format selection
(if A1=spec)
; X.xxx; XXX.x; XXXX
XX.xx
A3
Entry of damping
(1=no damping)
1-60
10
A5
Appendix
Liquisys M CUM 223 / 253
76
Hauser