Coding
Field
Selection or range
Factory setting
Display
Info
R211
Switch function of
R2 (1) off or on
off
on
All settings are retained.
R212
Enter switch-on
point of contact.
9999 FNU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m3
200.0 %
entire measuring
range
Never set the switch-on point and
switch-off point to the same value.
(Comp. chap. 5.5.1.)
(Only the operating mode
selected in A1 appears.)
If two factory settings are given,
then the left-hand one is valid for
CUS 41, the right-hand one for
CUS 31.
R213
Enter switch-off
point of contact
9999 FNU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m3
200.0 %
entire measuring
range
Entering the switch-off point
selects either a max contact
(switch-off point < switch-on
point) or a min contact (switch-off
point > switch-on point) and
implements a constantly required
hysteresis (see Fig. 5.5).
If two factory settings are given,
then the left-hand one is valid for
CUS 41, the right-hand one for
CUS 31.
R214
Enter pickup delay
0 s
0 ... 2000 s
R215
Enter dropout
delay
0 s
0 ... 2000 s
R216
Enter alarm
threshold
(as an absolute
value)
9999 FNU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m3
200.0 %
entire measuring
range
When the alarm threshold is
exceeded/undershot, the
measuring transmitter issues an
alarm with an error message and
error current (note alarm delay).
When defining the min contact,
the alarm threshold must be set
to a lower value than the
switch-off point.
If two factory settings are given,
then the left-hand one is valid for
CUS 41, the right-hand one for
CUS 31.
The factory seFactory settings are printed in
bold
face;
base version does not include functions in
italic.
off
R211
Function
9999
FNU
R212
On value
9999
FNU
R213
Off value
0
s
R214
On Delay
0
s
R215
Off Delay
9999
FNU
R216
A.Thresh
Instrument configuration
Liquisys M CUM 223 / 253
36
Hauser