7.2
Troubleshooting using the error messages
Display and select error messages by pressing the MINUS key.
Error
no.
Display
Measures
Contact
Error current
Cleaning trigger
Fact
User
Fact
User
Fact
User
E001
EEPROM memory error
Switch instrument off and back on,
return instrument to your local
Hauser sales agency for repair
or replace instrument.
Load software compatible with hardware.
Load measuring parameter specific
instrument software.
yes
no
—
—*
E002
Instrument not calibrated,
calibration data invalid,
no user data invalid
(EEPROM error).
yes
no
—
—*
E003
Download error
Invalid configuration. Repeat download,
check optoscope.
yes
no
—
—*
E004
Instrument software
version not compatible
with module hardware
version
Load software compatible with hardware.
Load measuring parameter specific
instrument software.
yes
no
—
—*
E007
Transmitter malfunction
Software not compatible
with hardware (transmitter)
yes
no
—
—*
E008
Sensor or sensor
connection defective
Check sensor and sensor connection
(E+H Sevice)
yes
no
no
E026
Wiper error
Check wiper and test function using
manual control if necessary
yes
no
no
E045
Calibration aborted
Repeat calibration
no
no
—
—*
E055
Below main parameter
measuring range
Check measurement and connections;
check instrument and measuring cable
with simulator if necessary.
yes
no
no
E057
Main parameter
measuring range
exceeded
Check measurement, control and
connections
yes
no
no
E059
Below temperature
measuring range
yes
no
no
E061
Temperature measuring
range exceeded
yes
no
no
E063
Below current output
range 1
Check configuration
yes
no
no
E064
Current output range 1
exceeded
Check measured value and current
assignment
yes
no
no
E065
Below current output
range 2
yes
no
no
E066
Current output range 2
exceeded
yes
no
no
E067
Set point exceeded
controller 1
Check configuration.
yes
no
no
E068
Set point exceeded
controller 2
yes
no
no
E069
Set point exceeded
controller 3
Check configuration
yes
no
no
E070
Set point exceeded
controller 4
yes
no
no
UM253E07.CHP
Liquisys M CUM 223 / 253
Maintenance and troubleshooting