Dropout delay
setting
0 ... 2000 s
0 s
R225
Setting of alarm threshold
(as an absolute value)
100.0 °C
–5.0 ... +100.0 °C
R226
Entry of derivate
action time Tv
0.0...999.9 min
0.0 min
R235
Selection of
controller characteristic
inv = inverted
dir = direct
R236
Selection
freq = pulse frequency
Ien = pulse length
R237
Entry of minimum
ON time t
0.1 ... 5.0 s
on
0.3 s
R2310
Entry of max.
pulse frequency
60 ... 180 1/min
120 1/min
R239
Entry of
pulse interval
0.5...999.9 s
10.0 s
R238
Dropout delay
setting
0 ... 2000 s
0 s
R215
Setting of alarm threshold
(as an absolute value)
entire measuring range
9999 FNU; 9999 ppm (mg/l);
300.0 g/l; 200.0 %
R216
Entry of post-rinse time
0 ... 999 s
20 s
R255
Number of repeat cycles
0 ... 5
0
R256
Set interval between
two cleaning cycles
(pause time)
1 ... 7200 min
360 min
R257
Number of cleaning
cycles without
cleaning agents
0 ... 9
0
R259
Set minimum
pause time
1 ... R357 min
120 min
R258
Entry of concentration
entire measuring range
K7
Table status o.k.
no
yes;
K8
Order number
is displayed
S7
Serial number
is displayed
S8
Perform instrument test
Displ = display test
no
S10
Reset of instrument
to basic values
Sens = sensor data;
Facty = factory settings
no
S9