Coding
Field
Selection or range
Factory setting
Display
Info
C114
Calibration status
is displayed
o.k.
E xxx
C115
Store calibration
results
yes
no
new
If C114 = E xxx, then only no or
new
(Exception: calibration
warning E 84).
If new, return to C.
If yes/no, return to
“Measurement”.
C1 (2)
Selection of
calibration
3 Pt = Three-point
calibration (1)
Corr
=
Three-point
correction (2)
Edit = Edit
calibration (3)
Refl = Fitting
with reflection
effects (4)
1-Pt = Single-point
calibration (5)
Data =Calibration
data (6)
C121
Enter correct
concentration of
the third
calibration
solution
Current value from
C113
entire measuring
range
In case of an unknown probe
concentration, but with a definite
dilution (
1
/
10
;
1
/
3
; 1), the
laboratory value is to be entered .
C122
Calibration status
is displayed
o.k.
E xxx
C123
Store calibration
results
yes
no
new
If C122 = E xxx, then only no or
new
(Exception: calibration
warning E 84).
If new, return to C.
If yes/no, return to
“Measurement”.
Factory settings are printed in
bold
face;
base version does not include functions in
italic.
o.k.
C114
Status
E045
E045
C161 ... C163
20 %
50 %
E084
E084
200 %
500 %
Cancel Warning:
Cancel
Warning:
yes
C115
Store
Corr
C1
Calibrat
1000.0
C121
Concentr3
FNU
o.k.
C122
Status
yes
C123
Store
Instrument configuration
Liquisys M CUM 223 / 253
50
Hauser