Function
Switch off or on
on
R2 (2)
off;
R221
Entry of switch-on
temperature
100.0 °C
–5.0 ... +100.0 °C
R222
Entry of switch-off
temperature
100 °C
–5.0 ... +100.0 °C
Pickup delay
setting
0 ... 2000 s
0 s
R224
R223
Function
Switch off or on
on
R2 (3)
off;
R231
Entry of set value
entire measuring range
0 FNU; 0 ppm (mg/l);
0 g/l; 0 %
R232
Entry of
control gain Kp
0.01 ... 20.00
1.00
R233
Entry of integral action time
Tn (0.0 = no I component)
0.0 ... 999.9 min
0.0 min
R234
Function
Switch off or on
on
R2 (4)
off;
R241
Rinse time setting
3 ... 999 s
30 s
R242
Pause time setting
1 ... 7200 min
360 min
Set minimum
pause time
1 ... 3600 min
120 min
R243
R244
Limit contactor
configuration
Clean =
Chemoclean
(only with rel3 and rel4)
Timer
PID controller
LC °C =
T limit contactor
LC PV =
TU / TS
Limit contactor
R2 (5)
R2 (4)
R2 (3)
R2 (2)
R2 (1)
Function
Switch off or on
on
R2 (1)
off;
R211
Select contact
switch-on point
entire measuring range
9999 FNU; 9999 ppm (mg/l);
300.0 g/l; 200.0 %
R212
Select contact
switch-off point
entire measuring range
9999 FNU; 9999 ppm (mg/l);
300 g/l; 200 %
R213
Pickup delay
setting
0 ... 2000 s
0 s
R214
Function
Switch off or on
on
R2 (5)
off;
R251
Entry of pre-rinse time
0 ... 999 s
20 s
R253
Entry of cleaning time
0 ... 999 s
10 s
R254
Start pulse selecction
;
ext = external;
i+ext = iexternal;
i+stp = internal
suppressed by ext
int = internal
R252
Function group
C
ONCENTRATION
MEASUREMENT
K
Selection of concentration
curve for calibration of
display value
Curve
... 4
1
K1
Table option
selection
edit
read
K3
Set number of
value pairs
1 ... 10
1
K4
Value pair selection
1 ... number of
value pairs in
1
K4
K5
Entry of turbidity
value
entire measuring range
0 FNU / ppm (mg/l) /
g/l / %
K6
R
Function group
C
ONTACTS
Select contact to be
configuredl
Rel2; Rel3; Rel4
Rel1;
R1
Select table to be
edited
1 ... 4
1
K2
Function group
CHECK
P
P1
PCS alarm setting
(live-check)
/ 1h / 2h / 4h
off
Function group
SERVICE
S
Language selection
GER
ITA; FRA
ESP; NEL
ENG;
S1
Hold configuration
=
CAL =during calibration
Setup =during setup
none =no hold
s+c during setup
and calibration
S2
Entry hold dwell
period
0...999 s
10 s
S4
Manual hold
on
off;
S3
Entry of SW upgrade
release code
(Plus package)
0000...9999
0000
S5
Entry of SW upgrade
release code
Chemoclean
0000...9999
0000
S6
Hardware
version
HW version
E112
Serial number
is displayed
E113
Software
version
SW version
E121
Hardware
version
HW version
E122
Serial number
is displayed
E123
Software
version
SW version
E131
Hardware
version
HW version
E132
Serial number
is displayed
E133
Software
version
SW version
E141
Software
version
SW version
E111
Entry of address
HART:
... 15
PROFIBUS: 1 ...
0
126
I1
Hardware
version
HW version
E142
Serial number
is displayed
E143
Function group
E + H SERVICE
E
Function group
INTERFACE
I
Module selection
Rel = relay
MainB
= mainboard
Trans
= transmitter
Contr
= controller
E1(4)
E1(3)
E1(2)
E1(1)
Module name
is displayed
E134
Module name
is displayed
E144
Module name
is displayed
E114
Module name
is displayed
E124
Appendix
Liquisys M CUM 223 / 253
78