Diagnostics and troubleshooting
Liquiline CM44P
152
Hauser
No.
Message
Factory settings
Sensor
type
Tests or remedial action
S
D
F
198
Operating time
M
On
Off
C
Operating hours > 150 °C, measurement
can still take place
1.
Replace sensor.
2.
Change monitoring limit.
3.
Disable monitoring.
199
Operating time
M
On
Off
All except
U, Phot
Total operating hours
215
Simulation active
C
On
Off
All except
Phot
Simulation active
End simulation by changing to measuring
mode.
408
Calibration aborted M
Off
Off
P, C, O, I,
DI, Phot
Calibration aborted
500
Sensor calibration
M
On
Off
All
Calibration aborted, main measured
value varies
Reasons: sensor too old, sensor
occasionally dry, calibration value not
constant
1.
Check sensor.
2.
Check calibration solution.
501
Sensor calibration
M
On
Off
All except
U, Phot
Calibration aborted, temperature
measured value varies
Reasons: sensor too old, sensor
occasionally dry, temperature of
calibration solution not constant
1.
Check sensor.
2.
Regulate calibration solution
temperature.
505
Sensor calibration
M
On
Off
P, O, I, DI
Max. zero point warning, measurement
can still take place
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1.
Check or replace sensor.
2.
Check or replace calibration
solution.
3.
Repeat calibration.
507
Sensor calibration
M
On
Off
P, O, I, DI
Min. zero point warning, measurement
can still take place
Possible reasons: sensor old or defective,
reference blocked, calibration solution
too old or contaminated
1.
Check or replace sensor.
2.
Check or replace calibration
solution.
3.
Repeat calibration.