Liquiline CM44P
Diagnostics and troubleshooting
Hauser
157
No.
Message
Factory settings
Sensor
type
Tests or remedial action
S
D
F
904
Process check
alarm
F
On
On
All except
Phot
Stagnating measuring signal
Reasons: sensor in air, sensor fouling,
incorrect flow to sensor, sensor defective
1.
Check electrode system.
2.
Check sensor.
3.
Restart the device.
914
USP/ EP alarm
M
On
Off
C
USP limit values exceeded
‣
Check process.
915
USP / EP warning
M
On
Off
C
916
EasyCal certificate
M
On
Off
Phot
EasyCal certificate expired
1. Send EasyCal for recertification and
enter new recertification date in Setup /
Photometer / calibration
2. If not used deactivate EasyCal
934
Process temp. high S
Off
Off
N, S, U
Process temperature high
1.
Do not increase process
temperature.
2.
Check electrode system.
3.
Change sensor type.
935
Process temp. low
S
Off
Off
N, S, U
Process temperature low
1.
Do not lower process temperature.
2.
Check electrode system.
3.
Change sensor type.
942
Process value
S
Off
Off
N, P, U
Process value high
1.
Do not increase process value.
2.
Check electrode system.
3.
Change sensor type.
943
Process value
S
Off
Off
N, P, U
Process value low
1.
Do not decrease process value.
2.
Check electrode system.
3.
Change sensor type.
944
Sensor range
S
On
Off
S, U
Measurement at periphery of sensor
dynamic range
Reasons: changes in process to a higher
or lower measuring range
1.
Check application.
2.
Use sensor that suits the
measuring range of the
application.
945
pH value high
S
On
Off
DI
Warning maximum pH value exceeded
1.
Check application.
2.
Check pH sensor.
946
pH value low
S
On
Off
DI
Warning minimum pH value not reached.
Potential leaking of gaseous chlorine!
1.
Check application.
2.
Check pH sensor.