![Endress+Hauser Liquiline CM44P Operating Instructions Manual Download Page 40](http://html.mh-extra.com/html/endress-hauser/liquiline-cm44p/liquiline-cm44p_operating-instructions-manual_2407399040.webp)
Electrical connection
Liquiline CM44P
40
Hauser
Sensor
Cable color
PEM terminal
Assignment
OUSBT66
BN
P+
Lamp v
BN
S+
Recording lamp v
BK
P-
Lamp voltage -
BK
S-
Recording lamp voltage -
RD
A (1)
OG
C(1)
Sensor -
TP
SH (1)
Shield
1)
OUSAF12
2)
OUSAF11
Memosens connection via M12 plug-in connection (field device only)
85
86
85
1
2
86
97
88
87
98
Sensor 1
PK
GY
GN
YE
BN
WH
1
A0018019
47
M12 plug-in connection (e.g.
at sensor module)
1
Sensor cable with M12
connector
6
4
1
NC
2
3
5
7
4
6
2
NC
1
7
5
3
A0018021
48
M12 assignment Top:
socket Bottom: connector
(top view in each case)
1
PK (24 V)
2
GY (Ground 24 V)
3
BN (3 V)
4
WH (Ground 3 V)
5
GN (Memosens)
6
YE (Memosens)
7,
NC
Not connected
Device versions with a pre-installed
M12 socket are ready-wired upon
delivery.
Version without a pre-installed M12
socket
1.
Insert an M12 socket
(accessory) into a suitable
opening in the base of the
housing.
2.
Connect the cable to a
Memosens terminal as per the
wiring diagram.
Connecting the sensor
‣
Plug the connector of the sensor
cable (→ 47item 1) directly into
the M12 socket.
Please note the following:
• The internal device wiring is always
the same regardless of what kind of
sensor you connect to the M12
socket (plug&play).
• The signal or power supply cables
are assigned in the sensor head in
such a way that the PK and GY
power supply cables are either used
(e.g. optical sensors) or not (e.g. pH
or ORP sensors).
6.4
Connecting additional inputs, outputs or relays
L
WARNING
Module not covered
No shock protection. Danger of electric shock!
‣
Change or extend the hardware: always fill the slots from left to right. Do not leave any
gaps.
‣
If all of the slots are not occupied: always insert a dummy cover or end cover in the slot
to the right of the last module(→ 2, 10). This ensures that the unit is shock-
protected.
‣
Always ensure shock protection is guaranteed particularly in the case of relay modules
(2R, 4R, AOR).
The terminal strip (cabinet unit) is used to connect the cable shields.