Diagnostics and troubleshooting
Hauser
147
No.
Message
Factory settings
Sensor
type
Tests or remedial action
S
D
F
140
Sensor check
F
On
Off
O
Sensor range errors
‣
Contact the Service Department.
141
Polarization
F
On
Off
C (cond.)
Polarization warning
The measured value is corrupted at high
conductivity levels.
‣
Use a sensor with a larger cell
constant.
142
Sensor signal
F
On
Off
C
Reasons: sensor in air, sensor defective
1.
Check installation.
2.
Replace sensor.
143
Sensor check
F
On
Off
C
Sensor self-test error
1.
Replace sensor.
2.
Contact the Service Department.
144
Conductivity range
S
Off
On
C
Conductivity outside measuring range
‣
Use a sensor with a suitable cell
constant.
146
Sensor temperature S
Off
Off
C, N, T, S
Temperature outside specification
1.
Check the temperature.
2.
Check electrode system.
3.
Replace sensor type.
147
Sensor check
F
On
On
C (ind.)
Coil transmission current too high
Reasons: transmission coil short-circuit,
inductance too low
1.
Replace sensor.
2.
Contact the Service Department.
148
Sensor check
F
On
On
C (ind.)
Reasons: transmission coil interrupted,
inductance too high
1.
Replace sensor.
2.
Contact the Service Department.
149
Sensor LED
F
On
On
T
Sensor LED error
1.
Replace sensor.
2.
Contact the Service Department.
151
Sensor buildup
F
On
On
T
Buildup, high degree of contamination
1.
Clean sensor.
2.
Replace sensor.
3.
Contact the Service Department.
152
Sensor data invalid M
Off
Off
C (ind.)
No calibration data
‣
Perform air set calibration.
153
Sensor defective
F
On
On
N, T, S,
Phot
Sensor lamp defective
Reasons: aging, end of operating life,
mechanical disturbance/vibration
1.
Replace sensor.
2.
Contact the Service Department.
154
Sensor data invalid M
Off
Off
C
Factory calibration is used
‣
Calibrate.