Diagnostics and troubleshooting
Liquiline CM44P
144
Hauser
No.
Message
Factory settings
Sensor
type
Tests or remedial action
S
D
F
005
Sensor data invalid F
On
On
All
1.
Check the firmware compatibility
of the sensor and transmitter or
load suitable firmware
2.
Set the sensor to the factory
settings, disconnect the sensor
and reconnect it.
3.
Update transmitter data
4.
Replace sensor.
010
Sensor scanning
F
Off
On
All
‣
Wait for initialization to be finished.
012
Writing data failed F
On
On
All
1.
Repeat write process.
2.
Replace sensor.
013
Sensor type wrong
F
On
On
All
Sensor does not suit the device
configuration or device configuration
must be changed to new type of sensor
1.
Change to a sensor of the type
that is configured.
2.
Adapt the device configuration to
the connected sensor.
018
Sensor not ready
F
On
On
All
Sensor communication blocked
1.
Sensor fails tag check. Replace.
2.
Internal software error. Contact
the Service Department.
022
Temperature
sensor
F
On
On
P, C, O, I,
DI
Temperature sensor defective
‣
Replace sensor.
061
Sensor electronic
F
On
On
All
Sensor electronics defective
‣
Replace sensor.
062
Sensor connection
F
On
On
All
1.
Check sensor connection.
2.
Contact the Service Department.
081
Initialization
F
On
On
All
‣
Wait for initialization to be finished.
100
Sensor
communication
F
On
On
All
Sensor not communicating
1.
Check sensor connection.
2.
Check sensor connector.
3.
Contact the Service Department.
101
Sensor
incompatible
F
On
On
All
1.
Update sensor firmware
2.
Replace sensor.
3.
Contact the Service Department.
102
Calibration timer
M
On
Off
All
Calibration interval elapsed.
Measurement can still take place.
‣
Calibrate sensor.
103
Calibration timer
M
On
Off
All
Calibration interval will elapse soon.
Measurement can still take place.
‣
Calibrate sensor.
104
Calibration validity M
On
Off
All
Last calibration no longer valid.
Measurement can still take place.
‣
Calibrate sensor.
105
Calibration validity M
On
Off
All
Last calibration will lose its validity soon.
Measurement can still take place.
‣
Calibrate sensor.