Liquiline CM44P
Calibration
Hauser
135
CAL/Photometer/Measurement channel (2nd measurement channel/Reference channel)/Application
adjustment
Function
Options
Info
Manual offset
Manual offset
Selection
• Yes
• No
Factory setting
No
Manual offset
0.000 to 9999 AU
Factory setting
0.000 AU
Current meas value
Read only
Decide whether you want to use the currently
displayed measured value as the offset.
Use current
measurement value as
offset
Action
Factory default
Action
This resets all the settings to the factory
calibration dataset.