Commissioning
Liquiline CM44P
60
Hauser
9.2.2
Display behavior
Menu/Operation/Display
Function
Options
Info
Contrast
5 to 95 %
Factory setting
50 %
Adjust the screen settings to suit your working
environment.
Backlight
=
Automatic
The backlighting is switched off automatically
after a short time if a button is not pressed. It
switches back on again as soon as you press the
navigator button.
Backlight
=
On
The backlighting does not switch off
automatically.
Backlight
Selection
• On
• Off
• Automatic
Factory setting
Automatic
Screen rotation
Selection
• Manual
• Automatic
Factory setting
Manual
If
Automatic
is selected, the single
‐
channel
measured value display switches from one
channel to the next every second.
9.3
User definable screens
Menu/Operation/User definable screens
Function
Options
Info
Meas. screen 1 ... 6
You can create 6 measuring screens of your own
and give them a name. The functions are
identical for all 6 measuring screens.
Meas. screen
Selection
• On
• Off
Factory setting
Off
Once you have defined your own measuring
screen, you can switch it on here. You can find
the new screen under
User definable screens
.
Label
Customized text, 20
characters
Name of the measuring screen
Appears in the status bar of the display.
Number of lines
1 to 8
Factory setting
8
Specify the number of measured values
displayed.
Line 1 ... 8
User interface
Label
Specify the content of
Label
in the submenu of
each line.
Source of data
Selection
• None
• See list in "Info" column
Factory setting
None
‣
Select a source of data.
You can choose from the following:
• Sensor inputs
• Heartbeat diagnostics of sensor inputs
• Controller
• Current inputs
• Fieldbus signals
• Mathematical functions
• Binary inputs and outputs
• Current outputs
• Relay
• Measuring range switching
Measured value
Source of data
is an
input
Selection
Depends on the input
Factory setting
None
You can display different main, secondary and
raw measured values depending on the type of
input.
No options can be selected for outputs here.