Operation
Liquiline CM44P
80
Hauser
Advanced setup
Measuring channel
Menu/Setup/Inputs/Photometer/Extended setup/Measurement channel
Function
Options
Info
Damping
0 to 600 s
Factory setting
0 s
The damping causes a floating average curve of
the measured values over the time specified.
Calib. settings
Stability criteria
You define the permitted variation in the
measured value which may not be exceeded in a
certain timeframe during calibration. If the
permitted difference is exceeded, the calibration
will not be allowed and is canceled
automatically.
Delta signal
0.1 to 2.0 %
Factory setting
0.2 %
Permitted measured value variation during
calibration
Duration
0 to 100 s
Factory setting
10 s
Timeframe within which the permitted
measured value variation may not be exceeded
Calibration timer
Selection
• Off
• On
Factory setting
Off
Switches the function on or off
Calibration timer value 1 to 10 000 h
Factory setting
1000 h
Enter the time after which the timer should time
out. When this time expires, diagnostic message
Calibration timer
is displayed with code 102.
Calib. expiration time
Selection
• Off
• On
Factory setting
Off
The function checks whether the calibration of a
sensor is still valid. Example: you install a pre-
calibrated sensor. The function checks how long
ago the sensor was last calibrated. A diagnostic
message is displayed if the last calibration is
longer in the past than the prespecified warning
or alarm limit.
Calib. expiration
time
Warning and alarm limits mutually influence
each other's possible adjustment range. Range in
which the two limits must be: 1 to 24 months.
Generally, the following applies: alarm limit >
warning limit
Warning limit
Factory setting
11 months
Diagnostic message: 105
Calibration validity
Alarm limit
Factory setting
12 months
Diagnostic message: 104
Calibration validity