Commissioning
Proline Promag P 500 HART
112
Hauser
10.4.10 Configuring empty pipe detection
The
Empty pipe detection
submenu contains parameters that must be configured for the
configuration of empty pipe detection.
Navigation
"Setup" menu → Empty pipe detection
‣
Empty pipe detection
Empty pipe detection
New adjustment
Progress
Switch point empty pipe detection
Response time empty pipe detection
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Empty pipe detection
–
Switch empty pipe detection on
and off.
• Off
• On
Off
New adjustment
The
On
option is selected in
the
Empty pipe detection
parameter.
Select type of adjustment.
• Cancel
• Empty pipe adjust
• Full pipe adjust
Cancel
Progress
The
On
option is selected in
the
Empty pipe detection
parameter.
Shows the progress.
• Ok
• Busy
• Not ok
–
Switch point empty pipe detection
The
On
option is selected in
the
Empty pipe detection
parameter.
Enter hysteresis in %, below
this value the measuring tube
will detected as empty.
0 to 100 %
10 %
Response time empty pipe detection In the
Empty pipe detection
On
option is selected.
Enter the time before
diagnostic message S862 ''Pipe
empty'' is displayed for empty
pipe detection.
0 to 100 s
1 s
10.4.11 Configuring the HART input
Navigation
"Setup" menu → HART input
‣
HART input
Capture mode
Device ID