16.4 Output
Output signal
HART current output
Current output
4 to 20 mA HART
Current span
Can be set to: 4 to 20 mA (active/passive)
Ex-i, passive
Open-circuit voltage
DC 28.8 V (active)
Maximum input voltage
DC 30 V (passive)
Load
250 to 700 Ω
Resolution
0.38 µA
Damping
Adjustable: 0.07 to 999 s
Assignable measured
variables
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Corrected conductivity
• Temperature
• Electronic temperature
Current output 0/4 to 20 mA
Current output
0/4 to 20 mA
Maximum output values
22.5 mA
Current span
Can be set to:
• 4 to 20 mA (active)
• 0/4 to 20 mA (passive)
Ex-i, passive
Open-circuit voltage
DC 28.8 V (active)
Maximum input voltage
DC 30 V (passive)
Load
0 to 700 Ω
Resolution
0.38 µA
Damping
Adjustable: 0.07 to 999 s
Assignable measured
variables
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Corrected conductivity
• Temperature
• Electronic temperature