Proline Promag P 500 HART
Diagnostics and troubleshooting
Hauser
161
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
495
Diagnostic event
simulation
Deactivate simulation
C
Warning
496
Status input simulation
Deactivate simulation status input
C
Warning
502
CT activation/
deactivation failed
Follow the sequence of the custody
transfer activation/deactivation: First
authorized user login, then set the DIP
switch on the main electonic module
C
Warning
511
ISEM settings faulty
1. Check measuring period and
integration time
2. Check sensor properties
C
Alarm
520
I/O 1 to n hardware
configuration invalid
1. Check I/O hardware configuration
2. Replace wrong I/O module
3. Plug the module of double pulse
output on correct slot
F
Alarm
530
Electrode cleaning is
running
Turn off ECC
C
Warning
531
Empty pipe adjustment
faulty
Execute EPD adjustment
S
Warning
537
Configuration
1. Check IP addresses in network
2. Change IP address
F
Warning
540
Custody transfer mode
failed
1. Check electronic components
2. Check parameter settings
3. Deactivate custody transfer mode
4. Reactivate custody transfer mode
F
Alarm
543
Double pulse output
1. Check process
2. Check pulse output settings
S
Warning
593
Double pulse output
simulation
Deactivate simulation pulse output
C
Warning
594
Relay output simulation Deactivate simulation switch output
C
Warning
599
Custody transfer
logbook full
1. Deactivate custody transfer mode
2. Clear custody transfer logbook (all
30 entries)
3. Activate custody transfer mode
F
Warning
Diagnostic of process
803
Current loop
1. Check wiring
2. Change I/O module
F
Alarm
832
Electronic temperature
too high
Reduce ambient temperature
S
Warning
833
Electronic temperature
too low
Increase ambient temperature
S
Warning
834
Process temperature too
high
Reduce process temperature
S
Warning
835
Process temperature too
low
Increase process temperature
S
Warning
842
Process limit
Low flow cut off active!
1. Check low flow cut off
configuration
S
Warning
882
Input signal
1. Check input configuration
2. Check external device or process
conditions
F
Alarm
937
Sensor symmetry
1. Eliminate external magnetic field
near sensor
2. Turn off diagnostic message
S
Warning