Proline Promag P 500 HART
Diagnostics and troubleshooting
Hauser
149
Power
I/O 2
I/O 3
I/O 3
1
2
3
4
5
3.
-
+
-
ESC
2.
Power
open
press
open
press
ESC
E
+
-
I/O 4
Power
I/O 2
I/O 3
1.
A0029689
1
Supply voltage
2
Device status
3
Not used
4
Communication
5
Service interface (CDI) active
1. Open the housing cover.
2. Remove the display module.
3. Fold open the terminal cover.
LED
Color
Meaning
1
Supply voltage
Off
Supply voltage is off or too low.
Green
Supply voltage is ok.
2
Device status
Off
Firmware error
Green
Device status is ok.
Flashing green
Device is not configured.
Flashing red
A diagnostic event with "Warning" diagnostic
behavior has occurred.
Red
A diagnostic event with "Alarm" diagnostic behavior
has occurred.
Flashing red/green
The device restarts.
3
Not used
–
–
4
Communication
Off
Communication not active.
White
Communication active.
5
Service interface (CDI)
Off
Not connected or no connection established.
Yellow
Connected and connection established.
Flashing yellow
Service interface active.
Proline 500
Different LEDs in the transmitter provide information on the device status.
Summary of Contents for Hart Proline Promag P 500