Commissioning
Proline Promag P 500 HART
114
Hauser
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Timeout
The
Burst network
option or
the
Master network
option is
selected in the
Capture mode
parameter.
Enter deadline for process
variable of external device.
If the waiting time is
exceeded, the diagnostic
message
F410 Data
transfer
is displayed.
1 to 120 s
5 s
Failure mode
In the
Capture mode
parameter, the
Burst network
option or
Master network
option is selected.
Define behavior if external
process variable is missed.
• Alarm
• Last valid value
• Defined value
Alarm
Failure value
The following conditions are
met:
• In the
Capture mode
parameter, the
Burst
network
option or
Master
network
option is selected.
• In the
Failure mode
parameter, the
Defined
value
option is selected.
Enter value to be used by the
device if input value from
external device is missing.
Signed floating-point
number
0
"Input" submenu
Navigation
"Expert" menu → Communication → HART input → Input
Parameter overview with brief description
Parameter
Description
User interface
Value
Signed floating-point number
Status
• Manual/Fixed
• Good
• Poor accuracy
• Bad
10.4.12 Configuring the relay output
The
Relay output
wizard guides the user systematically through all the parameters that
have to be set for configuring the relay output.
Navigation
"Setup" menu → Relay output 1 to n
‣
RelaisOutput 1 to n
Switch output function
Assign flow direction check
Assign limit
Assign diagnostic behavior
Assign status