WARNING:
DO NOT DIRECTLY SPRAY THE R100 with any disinfectants, or other liquids!
Français
ATTENTION
:
NE PAS pulvériser directement Le R100 avec les désinfectants ou d'autres liquides!
Important Note:
Spraying the unit would potentially allow the disinfectant to enter into the
Flow path, power and communications ports and damage these components, causing
it to fail.
Note Importante:
Pulvérisation de l'unité serait potentiellement permettre le désinfectant pour entrer
dans le Ports chemin d'écoulement, la puissance et de la communication et des
dommages à ces composants, l'amenant à coup sûr.
1.
Don a pair of clean or sterile gloves
Note
:
Gloved hands should be frequently sanitized with a disinfectant (e.g., 70% alcohol) throughout this
procedure.
2.
R100 Sanitization:
2.1
Before sanitizing the R100, disconnect the DC power supply cable, and insert the supplied Blower Exhaust
Port Plug (Blue), attach the Ethernet Port Cap (Attached Cap), and insert the USB Port Plug (small black
plug), to protect the receptacles, as shown in the following images:
Port Plugs
2.2
Sanitize the exterior of the R100, power supply and associated cables (if applicable), and IR Remote (if
applicable), with a wipe saturated with disinfectant. It is ideal to wipe the components with a primary
disinfectant (eg., Quaternary Ammonium) followed by a secondary disinfectant, preferably 70% IPA, or
Ethanol to remove disinfectant residues.