EMTEK, LLC - R100 USERS MANUAL / AUG2019.REV01
57
1.4
Initiate unit power as follows:
1.4.1 Plug in the DC power supply and connect it to the R100 Controller.
1.4.2 Press the unit power ON/OFF button to power up the R100.
1.4.3
Set the current Date/Time on the R100 (see Section 7.1.8).
1.5
R100 RAS Media Stage Adjustment:
1.5.1 Adjust the media stage (FIG A) by turning it clockwise to lower, or counter clockwise to raise
the stage, to achieve the desired stage height.
1.5.1.1
The stage should be lowered or raised to a point that the distance gauge, located on the
interior center surface of the R100 RAS inlet cover (FIG B), makes a mark on the
center of the surface of a test media plate (FIG C) that looks like FIG D. If the media
stage is too high, you will see a mark on the media surface similar to FIG E (media
crushed). If the media stage is too low, you will not see a mark on the media surface,
as in FIG F. It may take an additional test plate, and a little trial and error to achieve
the proper height for sampling on each testing day.
FIG A
Power ON/OFF
Power
Adapter Port