EMTEK, LLC - R100 USERS MANUAL / AUG2019.REV01
60
1.9 Aseptically remove the test plate as follows:
NOTE:
Gloved hands should be rinsed with secondary disinfectant and allowed to dry immediately
prior to performing these steps.
1.9.1 Remove the inlet cover by turning it counter clockwise to unlock it (Fig A). Hold the inlet
cover above the test plate with one hand (Fig B), and with the other hand replace the lid of the
test plate (Fig C). Be careful not to touch the inside of the lid or the agar surface. Do not
move your hand over the exposed surface of the test plate. Allow the lid to lead your hand
over the test plate as you replace it.
Fig A Fig B Fig C
1.9.2 Remove the test plate, with the lid in place (Fig A-Fig B), from the base plate holder and replace
the RAS inlet cover (Fig C).
Fig A Fig B Fig C
1.9.3 Secure the test plate lid to the test plate with tape.