b) Install a regulator on the clean air/percent known oxygen content gas
cylinder.
c) Attach a length of calibration tubing (PVC tubing, 3/16-in. ID, 5/16-in. OD)
from the regulator on the cylinder to the fitting on the IP filter assembly (part
number 00628-9000-0001).
d) Release the clean air/known percent oxygen specified calibration gas to the
sensor.
Note
If you need a long length of calibration tubing to reach the device, make
allowances for a delay in response time from the sensor while the clean air
travels the length of the calibration tubing.
e) Complete
through
.
f) Turn off the clean air (or percent oxygen specified calibration gas) when the
sensor is correctly zeroed.