background image

 
 
 
 
 
 
 
 
These instructions describe how to replace the IR Sensor in the H25-IR Infrared Gas Leak Detector.  
 
It is assumed that the user is familiar with the operation and menu system of the H25-IR. If necessary, refer to 
Instruction 3015-4342 or 3015-5286 (dependent on generation of model) for detailed operation and 
maintenance information. 
 

Items Required: 

 

Replacement IR Sensor Kit:* 

Model 

Bench Kit 

Solo Bench (450 mW)**  Solo Bench (540 mW)** 

CFC, HCFC, HFC & 
Halogen gases 

3015-4501 3015-5086 

R600 3015-4502 

3015-5089 

SF

6

 3015-4561 

3015-5087 

CO

2

 3015-4562 

3015-5090 

** Not for individual sale.  Used for ID purposes only. (i.e. To order replacement kit for your bench which is 
labeled model number 3015-5086, place order for kit number 3015-4501.) 

 

 

Medium Phillips head screwdriver 

 

Small flat blade screwdriver 

 

Wire cutter (for removing cable ties) 

 

Kit contains IR Sensor of gas being detected, cable ties (Qty. 2), and this instruction sheet 

 

Top Cover Removal 

1.  Remove both the front and rear bezels by first disengaging the plastic tabs on the right and left sides of 

the bezel, and then pulling the bezel straight out.  

2.  Remove the two top screws that secure the front panel. 

3.  Remove the four screws that secure the rear panel.  

4.  Remove all rear panel electrical connectors. 

5.  Drop down the rear panel far enough to allow the top cover to clear, being careful not to stress the 

cables attached to the panel. 

6.  Remove top cover by sliding it toward rear of instrument.  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 