Note
You do not need a conventional calibration cup to calibrate the sensor. Connect
calibration tubing (PVC tubing, 3/16-in. ID, 5/16-in. OD) directly to the fitting on the IP
filter assembly (part number 00628-9000-0001).
2.8.1
Calibrate using Field Communicator
Note
Emerson developed the Field Communicator guided setup configuration procedures in
this manual using Emerson AMS Trex Device Communicator. The menus are identical to
those found in other Field Communicators, but you navigate using touch screens rather
than fast keys. Refer to the manual for your handheld communication device for more
information.
Procedure
1.
WARNING
Explosions
Do not connect to the COMM terminals when an explosive atmosphere is present.
Connect the HART
®
communication leads from the Field Communicator HART
terminals to the COMM terminals on the terminal block of the transmitter.
A B
A. +COMM terminal
B. -COMM terminal
2. Establish communication between the transmitter and the Field Communicator.
3. On the
Home
screen, select
Configure
.
Configuration
Reference Manual
September 2019
00809-0100-4928
48