19. Select
Next
.
The
Service Reminder
screen is displayed if a service reminder is configured and
enabled.
20. Select
Next
to accept the service reminder date or enter another date.
for more information.
21. When the gas concentration reading stabilizes at or near the target gas
concentration level, shut off the target gas flow at the regulator.
22. Detach the calibration tubing from the regulator on the target gas source and from
the IP filter inlet on the bottom of the sensor.
2.9
Manual setup
Manual setup includes all available configuration settings. You may use it to change
specific settings configured during initial setup without using the
Guided Setup
menus.
You may also use it to configure advanced optional settings.
Note
Emerson developed the Field Communicator manual setup configuration procedures in
this manual using Emerson AMS Trex Device Communicator. The menus are identical to
those found in other Field Communicators, but are navigated using touch screens rather
than fast keys. Refer to the manual for your handheld communicator device for more
information.
Configuration
Reference Manual
September 2019
00809-0100-4928
70
Emerson.com/Rosemount
Summary of Contents for Rosemount 928
Page 1: ...Reference Manual 00809 0100 4928 Rev AE September 2019 Rosemount 928 Wireless Gas Monitor ...
Page 4: ...4 ...
Page 8: ...Contents Reference Manual September 2019 00809 0100 4928 viii Emerson com Rosemount ...
Page 106: ...Configuration Reference Manual September 2019 00809 0100 4928 106 Emerson com Rosemount ...
Page 178: ...Troubleshooting Reference Manual September 2019 00809 0100 4928 178 Emerson com Rosemount ...