b) Install a regulator on the clean air/known percent oxygen content gas
cylinder.
c) Attach a length of calibration tubing (PVC tubing, 3/16-in. OD, 5/16-in. OD)
from the regulator on the cylinder to the IP filter on the bottom of the sensor.
d) Release the clean air/known percent oxygen specified calibration gas to the
sensor.
Note
If you need a long length of calibration tubing to reach the device, make
allowances for a delay in response time from the sensor while the clean air
travels the length of the calibration tubing.
e) Perform
through
.
f) Turn off the clean air/known percent oxygen specified calibration gas when
the sensor is correctly zeroed.
Reference Manual
Configuration
00809-0100-4928
September 2019
Rosemount 928
61
Summary of Contents for Rosemount 928
Page 1: ...Reference Manual 00809 0100 4928 Rev AE September 2019 Rosemount 928 Wireless Gas Monitor ...
Page 4: ...4 ...
Page 8: ...Contents Reference Manual September 2019 00809 0100 4928 viii Emerson com Rosemount ...
Page 106: ...Configuration Reference Manual September 2019 00809 0100 4928 106 Emerson com Rosemount ...
Page 178: ...Troubleshooting Reference Manual September 2019 00809 0100 4928 178 Emerson com Rosemount ...