11. Select
OK
to accept the new zero.
12. On the
Calibrate Sensor
screen, enter a gas concentration level that corresponds to
the concentration of calibration gas that will be applied during calibration.
For oxygen, use 20.9 percent oxygen from clean air. This step may be performed
with surrounding air if no contaminants are present.
13. Select
OK
.
Reference Manual
Configuration
00809-0100-4928
September 2019
Rosemount 928
53