
7.5
Application challenges
7.5.1
Managing disturbance echoes
There are two general methods for managing disturbance echoes:
•
Set amplitude threshold to filter out weak disturbance echoes and noise.
•
Use the suppress false echoes function to manage strong disturbance echoes.
Amplitude threshold
The amplitude threshold is used to filter out noise and disturbing echoes from the product
surface echo. The transmitter uses certain criteria to decide which type of echo peak that
is detected. Only echoes above the amplitude threshold might be considered the product
surface. The amplitude threshold can either be set to a constant value, or split into
sections as defined by up to 10 anchor points.
If necessary, a customized amplitude threshold section can for instance be used to remove
the influence from the tank nozzle, or disturbances close to the tank bottom. Additionally,
it might be needed in areas where there are occasionally strong echoes present, for
instance due to wide mixer blades. Suppressing false echoes may not be sufficient in those
areas.
Note
Do not create a customized amplitude threshold section around echoes which are already
registered as false echoes.
Adjust the amplitude threshold
If necessary, the amplitude threshold can be increased if a disturbance echo is interpreted
as the product surface. Alternatively, a lower threshold may be required to handle weak
surface echoes (e.g. due to excessive foaming or turbulence).
Prerequisites
The amplitude threshold is set at factory for optimum performance and should not
normally need an adjustment. Before changing the factory default, study the position and
amplitude of the different echo peaks.
It is recommended to adjust thresholds using Rosemount Radar Master Plus.
Procedure
1. Select
Service Tools
→
Echo Curve
.
Service and troubleshooting
Reference Manual
September 2022
00809-0100-4418
94
Emerson.com/Rosemount