7.4.7
Restore to default settings
This function restores the transmitter to default settings (user configuration is
overwritten).
Prerequisites
Before restoring the transmitter to default settings, it is recommended to backup the
device configuration. The backup file can be used to restore configuration at a later stage.
Procedure
1. Select
Service Tools
→
Maintenance
→
Reset/Restore
.
2. Select
Restore Default Settings
and follow the on-screen instructions.
Related information
Save a backup file of the device configuration
7.4.8
Reset the Bluetooth key
Procedure
1. Select
Configure
→
(Manual Setup)
→
Device Setup
→
Security
.
2. Under
Bluetooth
, select
Reset Bluetooth Security
and follow the on-screen
instructions.
3. To confirm the reset, reconnect the device to the mobile app using the default key.
7.4.9
Use the simulation mode
This function can be used to simulate measurements.
Procedure
1. Select
Service Tools
→
Simulate
.
2. Under
Simulate Measurement Values
, select desired transmitter variable and follow
the on-screen instructions.
Related information
7.4.10
View input registers
Measured data is continuously stored in the input registers. By viewing the contents of the
input registers, expert users can check that the transmitter works properly.
Procedure
1. Select
Configure
→
(Manual Setup)
→
Level Setup
→
Advanced
→
Expert Options
→
Input Registers
.
Service and troubleshooting