Level is stuck in full tank
Figure 7-3: Symptom
A
B
C
D
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Disturbing object near the antenna.
Recommended actions
•
Use the suppress false echoes function to manage strong disturbance echoes.
•
Analyze the echo curve and check amplitude thresholds.
•
Increase the Upper Null Zone.
•
Remove the disturbing object.
•
Move the transmitter to another position.
Possible cause
Product build-up on the antenna.
Recommended actions
•
Clean the antenna.
Related information
Service and troubleshooting