Alarm mode close to tank bottom
Symptom
When the product surface is near the sloped tank bottom, the transmitter enters alarm
mode.
Figure 7-14: Symptom
A
B
C
D
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Reduction of projected surface area close to sloping tank bottom.
Recommended actions
•
Verify the tank geometry parameters are configured correctly (especially the
Reference Height and Bottom Offset).
•
If measurement in this region is not crucial, increase the Empty Tank Detection
Area.
•
Verify the Bottom echo visible when tank is empty parameter is disabled.
Related information
Reference height
Bottom offset
Empty tank handling
Enable bottom echo visible when tank is empty
Reference Manual
Service and troubleshooting
00809-0100-4418
September 2022
Rosemount 3408 Level Transmitter
83