6.
NEVER allow the slings to contact the electronics enclosure or the transducer
cabling. Damage to the meter may occur. If the slings do come in contact with the
electronics enclosure, then remove the two bolts holding the enclosure to its base
and temporarily remove the head from the meter during the lifting operation
(Remove the two bolts holding the enclosure to its base and unplug the cable from
the Acquisition Module. Two screws hold this cable in place.) Use a spreader-bar on
the slings to prevent contact with the electronics.
7.
Once the lifting operation is complete, reattach and secure the electronics cable to
J3 on the Acquisition Module, return the electronics enclosure to its original
position, replace the bolts, and secure the enclosure in place. Lifting the meter with
the upper enclosure installed but without the bolts installed, may cause the
electronics to fall and cause personal injury or electronics damage.
Incorrect sling attachment
Figure 2-16:
8.
NEVER apply shock loads to the meter. Always lift the meter gradually. If shock
loading ever occurs, the slings must be inspected per manufacturer's procedures
prior to being placed in any further service.
Mechanical installation
Installation manual
43