c. Click the Set Flow Range Limits button and enter a positive value for the Flow
Analysis Lower Velocity Range and the Upper Velocity Range Limits. When the
velocity is outside of the limit parameters, an alert is triggered. Click the Next
button to display the Method Complete page.
15.
Click the Service Tools tab to access the device alerts, variables, trends and
maintenance statuses or to edit the configuration parameters.
a. Click the Service Tools|Alerts tab. If an alert condition exists, the alert type and
description displays. Recommended actions are listed to assist you in a
resolution. After you resolve the alert condition, click the Acknowledge button to
clear the alert. Click Apply to write the changes to the meter. If no alert condition
is active, click OK to close the device window.
AMS Device Manager - Service Tools Alerts
Figure 4-17:
b. If you change the device configuration, a confirmation dialog displays and
prompts you to write the changes to the meter. Click Yes to write the changes to
the meter or click No to cancel pending changes.
Configuration
Installation manual
105