4.2.2
Display units
The Meter volume units displayed are either U.S. Customary or Metric. To modify the
Display Units, configure the Meter units system in the Field Setup Wizard — General Page
•
U.S. Customary volume unit selections are:
-
Cubic feet
-
Thousand cubic feet
•
Metric volume unit selections are:
-
Cubic meters
•
Display units preceded by a plus or minus sign indicate forward and reverse flow
direction, as shown in the table below.
•
The local display Flow rate time units are modifiable by selecting the drop-down
arrow and clicking the time unit in the list box
•
Valid flow rate time units selections are:
-
second
-
hour
-
day
4.2.3
Scroll delay
The Scroll Delay is the time interval for the selected display items to be shown on the Local
Display. The default scroll delay setting is five seconds. Click the spin box up or down arrow
to increase or decrease the length of time an item displays.
1.
Select Finish to write the configuration settings to the meter.
2.
Save the meter configuration file, collect a Maintenance log and Waveforms to
document the “As Left” settings.
4.3
Using AMS Device Manager to configure the
meter
This procedure assumes you have AMS Device Manager installed on the host computer
and have downloaded the latest Daniel Liquid Ultrasonic Meter Device Description (DD).
If not installed, click the link below to download the AMS device installation tool kit.
http://www.emerson.com/en-us/documents-and-drawings
4.3.1
Installing AMS Device description
1.
Use the link above to search for the Device Description (DD) for your Daniel 3410
Series Gas Ultrasonic Flow Meter.
2.
Use the Filter Results By categories to narrow-down your search.
Configuration
94
Gas Ultrasonic Flow Meter