11.
Configure the Component indexes and the C6+ split. This page is available for Daniel
Gas Ultrasonic meters and displayed only if View Gas Chromatograph checkbox was
selected on the Startup page and if the Gas Chromatograph Setup page was
previously displayed.
a. Click Next to continue to the AGA8 page.
12.
Configure the properties necessary for the AGA8 calculations.
This page is only displayed for Daniel Gas Ultrasonic meters if both temperature and
pressure are set to Live or Fixed and Base condition correction is selected on the
Startup Page. Configuration parameters include:
•
Calculations performed internally (by the meter) or Externally
•
AGA8 method - Gross Method 1, Gross method 2 or Detail
•
Base temperature and pressure
•
Specific gravity - reference temperature and pressure
•
Volumetric gross heating value and reference temperature
•
Molar density reference temperature and pressure
•
Flow Mass density, flow compressibility and Base compressibility
•
Gas composition inputs - components and mole percent
a. Click Next to continue to the Continuous Flow Analysis page, if View Continuous
Flow Analysis setup was selected on the Startup page.
13.
Configure the Continuous Flow Analysis (optional). This page is only displayed for
Daniel Gas Ultrasonic meters if both temperature and pressure are set to Live or
Fixed and Base condition correction is selected on the Startup Page . Configuration
parameters include:
a. Enable SOS comparison (requires AGA 8 Detail method).
b. Enable liquid detection and Profile factor limit.
c. Enable Blockage. Enter the percent for Symmetry, Cross-flow, Chords A to D
turbulence.
d. Enable Internal bore buildup.
a. Click Next to continue to the Alarm Limits page.
14.
Configure Alarm Limits for flow analysis, reverse flow and diagnostic chord:
a. Set low and high flow limits for flow analysis alarms.
b. Enable / Disable Reverse Flow alarm.
c. Set Volume limit and low flow limit for reverse flow alarm.
d. Enable co-located meter flow alarm.
•
Set Error Limit
e. Enable co-located meter speed of sound alarm.
•
Set Error Limit
f. Enable Diagnostic chord for co-located secondary head (Model 3416 only).
Configuration
90
Gas Ultrasonic Flow Meter