21.2. The salinity influence compensation
Salinity of the solution influences the oxygen solubility in water and has to be
taken into consideration in case of measurements in mg/l. 1 g/l of salinity
change causes about 0.5% of the oxygen concentration change. The
meter enables entering the salinity value in g/l and counts the change of the
oxygen concentration in mg/l.
21.2.1. Automatic introduction of the salinity value
The meter enables automatic introduction of measured solution salinity. In
order to do it:
- enter the conductivity measuring function by pressing the
button;
- measure salinity of the solution in g/l unit with conversion to NaCl (accurate
description in the section 17.1);
- after stabilisation of the salinity result enter the oxygen measuring mode by
pressing the
button;
- press the
button until the
(salinity) symbol displays in the lower
row of the LCD;
- press the
button shortly; In the upper row of the LCD the value of
the measured salinity in g/l will be displayed automatically. If after pressing
the button an
(error) symbol and next the previous value of salinity is
displayed it informs that during the conductivity measurement the unit has
not been changed for g/l of NaCl;
- return to the measuring mode by pressing the
button.
Summary of Contents for CX-401
Page 1: ...WATERPROOF MULTIFUNCTION METER CX 401 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CX 401 multifunction meter v6 0x...
Page 15: ...9 Pic 2...
Page 25: ...19 II pH MEASUREMENT...
Page 26: ...20 User s manual for CX 401 multifunction meter v6 0x...
Page 44: ...38 User s manual for CX 401 multifunction meter v6 0x...
Page 45: ...39 III CONDUCTIVITY AND SALINITY MEASUREMENT...
Page 46: ...40 User s manual for CX 401 multifunction meter v6 0x...
Page 71: ...65 IV OXYGEN CONCENTRATION MEASUREMENT...
Page 72: ...66 User s manual for CX 401 multifunction meter v6 0x...
Page 86: ...80 User s manual for CX 401 multifunction meter v6 0x...
Page 87: ...81 V ATMOSPHERIC PRESSURE MEASUREMENT...
Page 88: ...82 User s manual for CX 401 multifunction meter v6 0x...
Page 90: ...84 User s manual for CX 401 multifunction meter v6 0x...
Page 91: ...85 VI REDOX POTENTIAL AND TEMPERATURE MEASUREMENT...
Page 92: ...86 User s manual for CX 401 multifunction meter v6 0x...
Page 96: ...90 User s manual for CX 401 multifunction meter v6 0x...
Page 97: ...91 VII OTHER...
Page 98: ...92 User s manual for CX 401 multifunction meter v6 0x...
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......