- 25 -
Each of the calibration points has a different range for entering the pH
buffers’ values. This limitation enables the meter to recognise the buffer
solutions automatically. Table 1 contains the manufacturer’s settings of the
values of pH buffer solutions used for calibration, which may be changed
according to the ranges given in this table. The range of introducing these
changes is wide for each of the calibration points, what enables using buffer
solutions of pH values differing significantly from these set by the
manufacturer. In every case the meter will automatically detect the
introduced solution.
Table 1
Calibration point
Factory value
Range of changes
1
2,000
0,800
÷
2,100
2
4,000
3,900
÷
4,100
3
7,000
6,800
÷
7,100
4
9,000
8,900
÷
10,200
5
12,000
11,800
÷
14,000
The meter recognises only the pH buffers values detected during calibration.
The pH values stored in unused calibration points do not affect the calibration
results.
During next calibrations there is no need to perform the actions
described above, unless the buffer solutions are changed to buffers of
different values. The pH values introduced to the meter’s memory by the
user are stored in non-volatile memory.
The manufacturer gives information about the pH values of the solutions at
different temperatures. This data may be useful for calibrating the electrode
in temperature other than 20 oC by introducing the value of the buffer
corresponding to its value in this temperature to the meter’s memory.
There is no possibility to enter pH values in ranges other than those given in
the table 1.
Summary of Contents for CX-401
Page 1: ...WATERPROOF MULTIFUNCTION METER CX 401 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CX 401 multifunction meter v6 0x...
Page 15: ...9 Pic 2...
Page 25: ...19 II pH MEASUREMENT...
Page 26: ...20 User s manual for CX 401 multifunction meter v6 0x...
Page 44: ...38 User s manual for CX 401 multifunction meter v6 0x...
Page 45: ...39 III CONDUCTIVITY AND SALINITY MEASUREMENT...
Page 46: ...40 User s manual for CX 401 multifunction meter v6 0x...
Page 71: ...65 IV OXYGEN CONCENTRATION MEASUREMENT...
Page 72: ...66 User s manual for CX 401 multifunction meter v6 0x...
Page 86: ...80 User s manual for CX 401 multifunction meter v6 0x...
Page 87: ...81 V ATMOSPHERIC PRESSURE MEASUREMENT...
Page 88: ...82 User s manual for CX 401 multifunction meter v6 0x...
Page 90: ...84 User s manual for CX 401 multifunction meter v6 0x...
Page 91: ...85 VI REDOX POTENTIAL AND TEMPERATURE MEASUREMENT...
Page 92: ...86 User s manual for CX 401 multifunction meter v6 0x...
Page 96: ...90 User s manual for CX 401 multifunction meter v6 0x...
Page 97: ...91 VII OTHER...
Page 98: ...92 User s manual for CX 401 multifunction meter v6 0x...
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......