To calibrate:
-
place the sensor in the vessel with 0% saturation solution (saturated
sodium sulphite solution);
-
press and hold the
button till the moment of appearing the CAL
symbol in the bottom left corner of the display, the meter will enter the
calibration mode and automatically change the unit to % (if it was
previously set to measurement in mg/l O
2
);
-
when the reading stabilises, press the
button. The reading will blink,
what will inform about recording the calibration value; at the same time
the corrected reading (0%) will be displayed in the upper row of the LCD;
-
take the sensor out, rinse it accurately with distilled water and leave it
in the air;
-
when the reading stabilises, press the
button. The reading will blink,
what will inform about recording the calibration value; at the same time
the corrected reading (100%) will be displayed in the upper row of the
LCD;
-
return to the measuring mode by pressing the
button.
If after pressing the
button the meter is unable to detect the value of the
standard solution (0% or 100%), an
symbol displays for a moment
instead of the reading (Pic. 40). In such case it is necessary to check the
condition of the membrane in the sensor.
Pic. 40.
In case of choosing the sensor number, entering the calibration mode
and escaping it without making calibration, the stored characteristic
will be erased and a standard characteristic will be adopted.
Summary of Contents for CX-401
Page 1: ...WATERPROOF MULTIFUNCTION METER CX 401 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CX 401 multifunction meter v6 0x...
Page 15: ...9 Pic 2...
Page 25: ...19 II pH MEASUREMENT...
Page 26: ...20 User s manual for CX 401 multifunction meter v6 0x...
Page 44: ...38 User s manual for CX 401 multifunction meter v6 0x...
Page 45: ...39 III CONDUCTIVITY AND SALINITY MEASUREMENT...
Page 46: ...40 User s manual for CX 401 multifunction meter v6 0x...
Page 71: ...65 IV OXYGEN CONCENTRATION MEASUREMENT...
Page 72: ...66 User s manual for CX 401 multifunction meter v6 0x...
Page 86: ...80 User s manual for CX 401 multifunction meter v6 0x...
Page 87: ...81 V ATMOSPHERIC PRESSURE MEASUREMENT...
Page 88: ...82 User s manual for CX 401 multifunction meter v6 0x...
Page 90: ...84 User s manual for CX 401 multifunction meter v6 0x...
Page 91: ...85 VI REDOX POTENTIAL AND TEMPERATURE MEASUREMENT...
Page 92: ...86 User s manual for CX 401 multifunction meter v6 0x...
Page 96: ...90 User s manual for CX 401 multifunction meter v6 0x...
Page 97: ...91 VII OTHER...
Page 98: ...92 User s manual for CX 401 multifunction meter v6 0x...
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......