- 57 -
16.3. Measurement with manual temperature compensation
Measurement with manual temperature compensation may be made in
stable work conditions, e.g., during measurements in laboratory, especially
with use of thermostat, or in case of the temperature probe damage.
Disconnecting of the temperature probe switches the meter to manual
temperature compensation.
In case of measurement with manual temperature compensation:
-
connect the conductivity cell to the Cond connector (Pic. 2);
-
disconnect the temperature probe;
-
switch the meter on with the
button;
-
choose the conductivity measuring unit (section 13.1);
-
if the conductivity cell was not calibrated, calibrate it according to the
chapter 15);
-
check or change the value of the reference temperature and the
α
coefficient;
-
immerse the conductivity cell in the standard solution and hold it at least
1 cm away from the bottom and walls of the vessel. The measuring cell
should be filled up with the standard solution and should not include any
air bubbles, the electrode’s surface should be evenly moistened*;
-
measure the temperature of the solution with a laboratory thermometer
and enter the value with the
,
buttons;
-
after stabilisation check the reading (Pic. 35).
Note: pressing the
,
buttons simultaneously sets the temperature
to the reference temperature value.
Pic. 35.
*
- air bubbles may be removed by moving the immersed cell. In order to facilitate moistening of the
electrodes, it is recommended to immerse the cell in distilled water with washing-up liquid, and then
wash it in distilled water..
Summary of Contents for CX-401
Page 1: ...WATERPROOF MULTIFUNCTION METER CX 401 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CX 401 multifunction meter v6 0x...
Page 15: ...9 Pic 2...
Page 25: ...19 II pH MEASUREMENT...
Page 26: ...20 User s manual for CX 401 multifunction meter v6 0x...
Page 44: ...38 User s manual for CX 401 multifunction meter v6 0x...
Page 45: ...39 III CONDUCTIVITY AND SALINITY MEASUREMENT...
Page 46: ...40 User s manual for CX 401 multifunction meter v6 0x...
Page 71: ...65 IV OXYGEN CONCENTRATION MEASUREMENT...
Page 72: ...66 User s manual for CX 401 multifunction meter v6 0x...
Page 86: ...80 User s manual for CX 401 multifunction meter v6 0x...
Page 87: ...81 V ATMOSPHERIC PRESSURE MEASUREMENT...
Page 88: ...82 User s manual for CX 401 multifunction meter v6 0x...
Page 90: ...84 User s manual for CX 401 multifunction meter v6 0x...
Page 91: ...85 VI REDOX POTENTIAL AND TEMPERATURE MEASUREMENT...
Page 92: ...86 User s manual for CX 401 multifunction meter v6 0x...
Page 96: ...90 User s manual for CX 401 multifunction meter v6 0x...
Page 97: ...91 VII OTHER...
Page 98: ...92 User s manual for CX 401 multifunction meter v6 0x...
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......