- 32 -
User‘s manual for CX-401 multifunction meter
v6.0x
9.1. Readout of the electrode parameters after calibration
When the pH electrode has been calibrated, it is possible to check its
parameters: zero offset and slope, in the pH measurement mode.
To check:
-
press the
button until the electrode number shows at the upper part
of the display (
,
or
symbol);
-
with the
,
buttons choose the number of the electrode that is to
be checked and press the
button shortly. The last calibration date will
be displayed in the following format: month – day – year (below, Pic. 16A);
press the
button again, a
symbol will be displayed in the lower
row, the zero offset will be visible in the upper row of the display (Pic. 16B);
A B
Pic. 16
Summary of Contents for CX-401
Page 1: ...WATERPROOF MULTIFUNCTION METER CX 401 USER S MANUAL...
Page 2: ......
Page 4: ......
Page 7: ...1 I INTRODUCTION...
Page 8: ...2 User s manual for CX 401 multifunction meter v6 0x...
Page 15: ...9 Pic 2...
Page 25: ...19 II pH MEASUREMENT...
Page 26: ...20 User s manual for CX 401 multifunction meter v6 0x...
Page 44: ...38 User s manual for CX 401 multifunction meter v6 0x...
Page 45: ...39 III CONDUCTIVITY AND SALINITY MEASUREMENT...
Page 46: ...40 User s manual for CX 401 multifunction meter v6 0x...
Page 71: ...65 IV OXYGEN CONCENTRATION MEASUREMENT...
Page 72: ...66 User s manual for CX 401 multifunction meter v6 0x...
Page 86: ...80 User s manual for CX 401 multifunction meter v6 0x...
Page 87: ...81 V ATMOSPHERIC PRESSURE MEASUREMENT...
Page 88: ...82 User s manual for CX 401 multifunction meter v6 0x...
Page 90: ...84 User s manual for CX 401 multifunction meter v6 0x...
Page 91: ...85 VI REDOX POTENTIAL AND TEMPERATURE MEASUREMENT...
Page 92: ...86 User s manual for CX 401 multifunction meter v6 0x...
Page 96: ...90 User s manual for CX 401 multifunction meter v6 0x...
Page 97: ...91 VII OTHER...
Page 98: ...92 User s manual for CX 401 multifunction meter v6 0x...
Page 116: ......
Page 117: ......
Page 118: ......
Page 119: ......