SECTION 6
NOZZLE MAINTENANCE AND PERFORMANCE
6.5 Using the Line Flush
NOTE:
You can still use the Cytocentrifuge mode during this time by selecting
Cyto.
11 Load 500 mL of deionized or distilled water into the line(s).
12 Press
Start
.
The instrument will pump approximately 400 mL of water
through the line(s). A status bar will indicate progress.
13 When the flush is complete, remove the remaining water.
14 Load at least 300 mL of the correct reagent for that position into the
line(s).
15 Press
Start
. The instrument will pump approximately 200 mL of the
reagent(s) through the line(s).
16 Reinstall nozzle(s). Make sure you reinstall the correct nozzle for each line.
17 Press
Continue
. The stainer cleans and primes the lines and returns to the
Main menu.