SECTION 4
OPERATING THE STAINER
4.1 Operating Instructions
NOTE:
You can access the Reagent Information menu by pressing the bottle icons on the
right side of the Main menu. Press
Change
to scan or manually enter reagent
information
.
NOTE:
Do not put residual reagent from a used bottle into a new bottle. This can lead to
an accumulation of residue on the slides and may be a source of contamination.
WARNING!
Reagents used in the instrument contain moderately hazardous chemicals that require
care in handling. Always use appropriate safety measures, including gloves and eye
protection, when handling reagents.
Replacing a Reagent Bottle
1
Remove the empty reagent bottle from the tray but do not disconnect the
dip tube.
2
Open the Reagent Information menu by pressing the reagent bottle icon
on the right side of the Main menu.
3
If you are using reagent tracking, scan the barcode, or manually enter the
reagent REF, expiration date, and lot number in the Using Reagent
Information Tracking menu (Section 3.1).
4
Open the new bottle and record the letter on the cap for future use such
as long-term storage.
5
Place the new bottle in the tray.
6 Select the desired reagent and press
Change.
7
Unscrew the cap and remove the dip tube from the empty bottle.
8
Insert the dip tube into the new reagent bottle and screw on the cap.
Emptying the Waste Container
The Reagent Level Detect function automatically monitors the waste level and indicates
when the waste container should be emptied. It is still necessary to check waste levels
visually to ensure the waste container does not overfill.
CAUTION:
Dispose of collected waste according to local statutes and safety requirements.
1
Unscrew the cap from the full waste container.
2
Discard the waste according to local regulations.
3
Reinstall the cap on the empty waste container.