SECTION 3
CONTROLLING AND CUSTOMIZING STAINER FUNCTIONS
3.1 System Setup Menu
Many software settings can be controlled from the System Setup menu, including:
•
Creating, editing, and deleting stain programs
•
Creating, editing, and deleting cytocentrifuge programs
•
Tracking reagent information
•
Managing user accounts
•
Enabling tracking features for slides, preventive maintenance, and reagents
•
Managing reagent and waste level sensing
•
Changing the display language
•
Viewing and exporting the system log
•
Changing beeper settings
•
Setting the date and time
•
Restoring default settings
Accessing the System Setup Menu
1 Press
System Information
from the Main menu.
2
Press
System Setup
.
Creating a Stain Program
1
From System Setup, press
Stain Programs
.
2
Press
Add
.
3
Press
Enter Name
.
4
Enter a program name in the Program Name field.
5
Adjust the settings (see Adjusting Stain Settings on the next page).
Information concerning stain setting options is found in Appendix D.
6
Press
Save
.