48
SECTION 4
OPERATING THE STAINER
4.1 Operating Instructions
Suggested Staining Protocol
NOTE:
Samples and slides should be prepared and fixed according to recommendations
in Appendix E or equivalent.
•
Hub Pattern Test (once per day).
•
If slide tracking is enabled, scan or enter slide information.
•
Load slides into the carousel. Use blocking slides if needed.
•
Place loaded carousel into the stainer and close the lid.
•
Check reagent and waste levels.
•
If slide tracking is not enabled, enter the number of slides on the Main menu.
•
Select or verify desired stain program.
•
Perform a stain cycle.
•
Unload the carousel.
Performing a Hub Pattern Test
Use the Hub Pattern Test to ensure the nozzles are clear of debris and spraying properly.
1
From the Maintenance menu, Select
Pattern Test.
2
Hold a sheet of white paper towel near the drive hub, squarely facing the
target nozzle.
3
Select the
corresponding prime button.
4
Check the quality of the pattern. If the pattern is not correct, refer to
Nozzle Maintenance and Performance (Section 6).
Figure 7: Correct Hub Pattern Test Result
Figure 8: Incorrect Hub Pattern Test Result
NOTE:
If the Hub Pattern Test result is incorrect, clean the nozzle orifice with the nozzle brush
provided in the Nozzle Maintenance Kit.