28 ElektroPhysik
Press the
ò
-key to select NON-FERROUS, i.e. the
eddy-current method. If none of the keys is pressed,
the gauge will automatically switch to automatic
mode after about 3 sec.
This mode can be used if the substrate type (steel or
non-ferrous metal) is not known.
Calibration for the automatic mode requires a
previous measurement on the uncoated substrate.
The display will then show FERR or NON-FERR.
For calibration and measurement, proceed as nor-
mal according to either 3.2.2 or 3.2.3 or 3.2.5.
For alternating measurements on steel and non-
ferrous substrates, calibration must be performed
on uncoated samples of both substrates.
Measurements can then be carried out immediately.
3.2.5 Calibration and measurement on shot
blasted surfaces
The physical nature of shot-blasted surfaces results
in coating thickness readings which are too high.
Measurement on shot blasted surfaces