ElektroPhysik 27
Note:
-
This will reactivate the default standard
calibration for use on even surfaces.
-
Even while a series of measurements is being
taken, foil calibration can be carried out as
often as necessary. The old calibration will
be overwritten; the ZERO calibration remains
in memory.
3.2.4 Calibration and measurement with
MiniTest 600 FN
The universal probe of MiniTest 600 FN uses both
the magnetic induction (measuring range: 0 ... 3mm
(0 ... 0.12") and the eddy-current principle
(measuring range: 0 ... 2 mm (0 ... 0.08")).
To select the measuring method, press ON. The word
FERROUS will flash on screen. Press the
ñ
-key to
select FERROUS, i.e. the magnetic induction
method.
Calibration and measurement with MiniTest 600 FN